Infrared
Moisture Meters/ Infrared Thickness Meters
Infrared
Moisture Meters
|
 |
 |
|
 |
|

|
IR-M1000 Series
Infrared Moisture Meters
|
|
|
|
| The
IR-M1000 Series is reflection system on-line moisture
meters utilizing the infrared absorption of moisture.
Converting capabilities are built into the compact designed
detector unit for easy installation and operation. Maximum
99 calibration curves can be stored into memory for many
applications. The unit can be used by itself or by combining
with personal computers as an analog output and a communication
function are provided with. A setting
display unit connectable with maximum 9 detector units
can set various kinds of setting items of the detector
unit and displays moisture value. |
|
|
Features
- Built-in
converting capabilities
The converting
capabilities are built into the small, lightweight
and compact detector unit.
- Key
and display function
The detector unit has operation
keys and digital display function and can be used
by the unit itself. In addition, the calibration and
the creation of calibration curves at site are easy.
- Measurement
with high-speed and high sensitivity
By high-performance CPU and
improved optical components, high speed (output renewal
cycle: 28ms) and high sensitive measurement can be
performed.
- Setting
display unit
The setting display unit has a front
panel of 96 x 96 mm and can connect with up to 9 sets
of the detector unit. The communications interface
between the setting display unit and the detector
unit is by RS-485.
- Protective
structure conforming to IP-65
The enclosure of the detector
unit and the front panel of the setting display unit
conform to IP-65 for severe atmospheric condition.
- CE
Models complying with CE are available.
- Explosion-proof
construction
Models complying with [ExdIIBT4] is
available.
|
|
Models
|
Communications
interfaces
|
Measuring
objects
|
|
Universal
type
|
High
moisture type
|
Micro
moisture type
|
| RS-485
(standard) |
IR-M1100S
|
IR-M1200S
|
IR-M1300S
|
| RS-232C
|
IR-M1100R
|
IR-M1200R
|
IR-M1300R
|
| RS-422A
|
IR-M1100A
|
IR-M1200A
|
IR-M1300A
|
- Explosion-proof type
detector unit
|
Communications
interfaces
|
Exposion-proof
type
|
| With
display |
Without
display |
With
display For small diameter |
| RS-485
(standard) |
IR-M1 00SD
|
IR-M1 00SE
|
IR-M1 00SF
|
| RS-232C
|
IR-M1 00RD
|
IR-M1 00RE
|
IR-M1 00RF
|
|
Models
|
Communications
interfaces
|
| IR-GMEG1S
|
RS-485
|
| IR-GMEG1R
|
RS-232C
|
| IR-GMEG1A
|
RS-422A
|
Specifications
| Measuring
system |
Infrared
reflection type 3-wavelength system
(2-wavelength system for micro moisture type) |
| Measuring
distance |
300
mm (Distance of 200 to 400mm is possible.) |
| Measuring
area |
About
50 x 50mm (at measuring distance of 300mm)
30 x 30mm for small diameter
type
(30 x 30mm/measuring distance
200mm for micro moisture type) |
| Moisture
output |
4
to 20mADC (Load resistance: less than 500 )
Correction input: 4 to 20mADC (option)
(For correction of sample temperature,
measuring distance, etc.) |
| Communications
output |
RS-485,
RS-232C or RS-422A (to be specified)
RS-485 for connection with the setting display unit |
| Output
renewal cycle |
28ms |
| Display |
LED 5-digit
Calibration curve number - LED 2-digit |
| Setting |
By
keys or communications |
| Calibration
curve |
Linear
to cubic expression or broken line |
| Numbers
of calibration curve |
99
curves |
| Calibration
curve correction |
Zero
and span correction |
| Calibration
curve creation |
Regression
calculation |
| Calibration |
By
a checking plate |
| Smoothing |
0
to 99 seconds |
| Working
temperature range |
0
to 50°C(Air is necessary for more than 45°C.)
|
| Rated
supply voltage |
24VDC |
| Power
consumption |
Maximum
36VA |
| Connection |
Terminals
connection |
| Casing |
Aluminum
casing, drip-proof structure (conforming to IEC529
IP65) |
| Mounting |
Suspension
system using four M8 bolts |
| Weight |
About
4kg |
| Accessory |
Power
supply unit IR-WEP |
* Models complying with explosion-proof
construction [ExdIIBT4] are available.
| Detector
unit input |
RS-485,
connectable with up to 9 detector units |
| Analog
output |
4
to 20mADC (Load resistance: less than 500 ),
2 outputs |
| Communication
output |
RS-232C,
RS-485 or RS-422A (to be specified) |
| Output
renewal cycle |
28ms
x (quantity of detector unit) |
| Display |
Data
- LED 5-digit
Calibration curve number - LED 2-digit
Detector unit number - LED 1-digit |
| External
setting |
Setting
of detector unit number, calibration curve number,
calibration, preset output and hold by external
contact |
| Alarm |
High/low
alarms, Contact 2 outputs (H-C-L) |
| Working
ambient temperatur |
0
to 50°C |
| Rated
supply voltage |
100
to 240VAC 50/60Hz |
| Power
consumption |
Maximum
15VA |
| Casing |
ABS
resin, Front panel - Drip-proof structure (conforming
to IP-65) |
| Weight |
About
0.6kg |
| Output
voltage |
24VDC |
| Output
current |
2.1A |
| Working
ambient temperature |
0
to 50°C |
| Supply
voltage |
100
to 120VAC, 200 to 240VAC, switching type, 47 to
450Hz |
| Power
consumption |
Maximum
160VA |
| Weight |
About 380g |
| Analog
output
|
4
to 20mADC (Load resistance: less than 500 ),1
output
(1) Output accuracy - ±0.5%of full scale
(2) Stability - ±3% of full scale under EMC
test environment |
| Rated
power supply |
24VDC,
allowable voltage fluctuation - ±10% of rated
value |
| Power
consumption |
About
10VA
<Standard>
EN55011 group 1 class A, EN50082-2 (industrial environment)
<Directive>
89/336/EEC, 92/31/EEC (amendment), 93/68/EEC (amendment)
|
- Software package IR-WEF1(for
IR-M1000, IR-M2000)
| Contents
|
Data logging, trend display, data
storage and calibration curve creation
Up to 9 detector unit s are connectable. An electronic
balance is connectable on the creation of calibration
curve. |
| Environment |
DOS/V PC |
| OS |
Windows
95, 98, 2000 Contact input board (option): IBX-2131 (ISA bus)
|
Page Top
BACK
|
|
|
|