The IR-M2000 Series is
fiber optic on-line moisture meters utilizing the infrared
absorption of moisture. The sensing part is a fiber
optics and can be installed at narrow space and high
temperature environment.
Converting capabilities
are built into the compact designed detector unit for
easy installation and operation. Maximum 99 calibration
curves can be stored into memory for many applications.
The unit can be used by itself or by combining with
personal computers as an analog output and a communication
function are provided with.
A setting display unit
connectable with maximum 9 detector units can set various
kinds of setting items of the detector unit and displays
moisture value.
|
|
Models
|
Measuring
objects
|
Communications
interfaces
|
Applicable
fiber optics
|
|
Universal
types
|
High
moisture types
|
|
IR-M2100S
|
IR-M2200S
|
RS-485
(standard)
|
Reflection
type
|
|
IR-M2100R
|
IR-M2200R
|
RS-232C
|
|
IR-M2100A
|
IR-M2200A
|
RS-422A
|
|
IR-M2150S
|
IR-M2250S
|
RS-485
(standard)
|
Transparent
type
|
|
IR-M2150R
|
IR-M2250R
|
RS-232C
|
|
IR-M2150A
|
IR-M2250A
|
RS-422A
|
|
Models
|
Types
|
|
IR-WCRN
|
Reflection
type
|
|
IR-WCRE
|
Reflection
type with lens
|
|
IR-WCT
|
Transparent
type*
|
* For liquid, a measuring
cell for liquid is separately necessary.
|
Models
|
Communications
interfaces
|
|
IR-GMEG1S
|
RS-485
|
|
IR-GMEG1R
|
RS-232C
|
|
IR-GMEG1A
|
RS-422A
|
Specifications
| Measuring
system |
Infrared
reflection type 3-wavelength system |
| Moisture
output |
4
to 20mADC (Load resistance: less than 500 )
|
| Correction
input |
4 to 20mADC (option)
(For correction of sample temperature, measuring
distance, etc.)
|
| Communications
output |
RS-485,
RS-232C or RS-422A (to be specified)
RS-485 for connection with the setting display unit |
| Output
renewal cycle |
28ms |
| Display |
LED
5-digit
Calibration curve number - LED 2-digit |
| Setting |
By
keys or communications |
| Calibration
curve |
Linear
to cubic expression or broken line |
| Numbers
of calibration curve |
99
curves |
| Calibration
curve correction |
Zero
and span correction |
| Calibration
curve creation |
Regression
calculation |
| Calibration |
By
a checking plate |
| Smoothing |
0
to 99 seconds |
| Working
temperature range |
0
to 50 °C(Air is necessary for more than 45°C.)
|
| Rated
supply voltage |
24VDC |
| Power
consumption |
Maximum
36VA |
| Connection |
Fiber optics - Connector
connection
Power supply/signal line -Terminal connection
|
| Casing |
Aluminum
casing, drip-proof structure (conforming to IEC529
IP65) |
| Weight |
About
4kg |
| Accessory |
Power
supply unit IR-WEP |
| Measuring
distance and measuring diameter |
Reflection
type - 20mm
at 15mm, 50mm
at 50mm
Reflection type with lens - 25mm
at 25mm, 40mm
at 100mm |
| Fiber
length |
Reflection type,
Reflection type with lens - 1.5m (up to 10m)
Transparent type - 2m (up to 20m)
|
| Allowable
bending radius |
Reflection type,
Reflection type with lens - R100mm
Transparent type - R250mm
|
| Working
temperature range |
0
to 150°C |
| Fiber
protection |
Stainless
flexible tube |
| Purge
air flow |
Reflection type -
5 to 20Nl/min
Reflection type with lens - Without air purge
function
|
| Detector
unit input |
RS-485,
connectable with up to 9 detector units |
| Analog
output |
4
to 20mADC (Load resistance: less than 500 ),
2 outputs |
| Output
renewal cycle |
28ms
x (quantity of detector unit) |
| Display |
Data
- LED 5-digit
Calibration curve number - LED 2-digit
Detector unit number - LED 1-digit |
| External
setting |
Setting of detector
unit number, calibration curve number, calibration,
preset output and hold by external contact
|
| Alarm |
High/low
alarms, Contact 2 outputs (H-C-L) |
| Working
ambient temperature |
0
to 50°C |
| Rated
supply voltage |
100
to 240VAC 50/60Hz |
| Power
consumption |
Maximum
15VA |
| Casing |
ABS
resin, Front panel - Drip-proof structure (conforming
to IP-65) |
| Weight |
About
0.6kg |
| Output
voltage |
24VDC
|
| Output
current |
2.1A |
| Working
ambient temperature |
0
to 50°C |
| Supply
voltage |
100
to 120VAC, 200 to 240VAC, switching type, 47 to
450Hz |
| Power
consumption |
Maximum
160VA |
| Weight |
About
380g |
- Package software Contents
| Contents
|
Data
logging, trend display, data storage and calibration
curve creation
Up to 9 detector unit s are connectable.An electronic
balance is connectable on the creation of calibration
curve. |
| Environment |
IBM
PC/AT compatible |
| OS |
Windows
95 Contact input board (option): IBX-2131 (ISA bus)
|
Page Top
BACK
|