IRM-V Series Infrared Constituent Meter
CHINO Japanese Page
Home Products Corporate Infomation Downloads Request Catalog

Infrared Moisture Meters/ Infrared Thickness Meters
Infrared Constituent Meter

IRM-V Series Infrared Constituent Meter

 

 

The IRM-V series is near-infrared sensors utilizing the near-infrared absorption for measurements of moisture, coating thickness, film thickness and organic matters. By using narrow band filters (maximum 6 filters) corresponding to absorption spectral of OH group and CH group in the near-infrared band, it can simultaneously measure 1 to 3 constituents with various computation processing.
IRM-V
Features
  • Simultaneous measurement up to 3 constituents
    (moisture/thickness of film, moisture/albuminoid/lipid of powder milk, etc.)
  • No need to change a calibration curve of specific objects, calibration curve free moisture meter
  • Easy-to-operation
  • Flexible system configuration
  • Highly graded data processing function

Models

  • Constituent meter detector unit
Measuring
wavelength
Mirror system Fiber optic system Mirror system
Special spot size
Explosion-proof type
Mirror system Fiber optic system
3-wavelength
General type
IRM82112 IRM82212 IRM82912 IRM82512 IRM82612
3-wavelength
High moisture type
IRM82122 IRM82222 IRM82922 IRM82522 IRM82622
3-wavelength * IRM821A2 IRM822A2 IRM829A2 IRM825A2 IRM826A2
5-wavelength * IRM821B2 IRM822B2 IRM829B2 IRM825B2 IRM826B2
6-wavelength * IRM821C2 IRM822C2 IRM829C2 IRM825C2 IRM826C2
In each model, 4 types of "1 analog output + 2 alarm outputs", "2 analog outputs + 1 alarm output", "3 analog outputs" and "1 analog output + 1 Pt input" are available in optional [detector input/output specifications].

* Specify wavelength.
  • Trace moisture meter detector unit (Mirror system)

Measuring wavelength

Models

2-wavelength, trace moisture IRM82332
3-wavelength, trace moisture * IRM823A2
5-wavelength, trace moisture * IRM823B2
6-wavelength, trace moisture * IRM823C2
In each model, 4 types of "1 analog output + 2 alarm outputs", "2 analog outputs + 1 alarm output", "3 analog outputs" and "1 analog output + 1 Pt input" are available in optional [detector input/output specifications].

* Specify wavelength.

  • Thickness meter (Film thickness/coating thickness)
Measuring
wavelength
Mirror system Fiber optic system Mirror system
Special spot size
Explosion-proof type
Mirror system Fiber optic system
2-wavelength
Film thickness
IRM82132 IRM82232 IRM82932 IRM82532 IRM82632
3-wavelength
Film thickness
IRM82142 IRM82242 IRM82942 IRM82542 IRM82642
In each model, 4 types of "1 analog output + 2 alarm outputs", "2 analog outputs + 1 alarm output", "3 analog outputs" and "1 analog output + 1 Pt input" are available in optional [detector input/output specifications].
  • Thickness meter (Mirror system)

Measuring wavelength

Near-infrared ultra-thin thickness meter

Infrared ultra-thin thickness meter

Projection/ receiving separated type film thickness meter

Explosion-proof type infrared ultra-thin thickness meter

2-wavelength Thickness IRM82732 IRM82A32 IRM828312 IRM82B32
3-wavelength Thickness IRM82742 IRM82A42 IRM828412 IRM82B42
3-wavelength Thickness * IRM827A2 IRM82AA2 IRM828A12 IRM82BA2
5-wavelength Thickness * IRM827B2 IRM82AB2 IRM828B12 IRM82BB2
6-wavelength Thickness * IRM827C2 IRM82AC2 IRM828C12 IRM82BC2
In each model, 4 types of "1 analog output + 2 alarm outputs", "2 analog outputs + 1 alarm output", "3 analog outputs" and "1 analog output + 1 Pt input" are available in optional [detector input/output specifications]. (1 analog output is only available in the projection/receiving separated type.)

* Specify wavelength.
  • Fiber optics

Models

Types

IR-WCRN Reflection type fiber optics
IR-WCRE Reflection type fiber optics with lens

Specifications

  • Constituent meter detector unit
Measuring system Infrared reflection type multiple wavelength system
Optics Mirror system, Fiber optics system, Mirror system for special spot diameter
Wavelength 2, 3, 5, 6-wavelength
Constituent 1, 2, 3 constituents
Measuring distance Mirror system … 300mm (200 to 400mm possible)
Fiber optics system … Shown separately
Measuring area
Mirror system 50 x 50mm (at 300mm measuring distance)
Special spot diameter 50 x 100mm, phi50mm
Fiber optics system … Shown separately
Input/output terminals Select from the followings:
1. 1 analog output + 2 alarm outputs
2. 2 analog outputs + 1 alarm output
3. 3 analog outputs
4. 1 analog output +1 Pt100 input (for sample temperature correction)
Analog output 4 to 20mADC or 0 to 20mADC (load resistance: less than 500ohm)
Analog output accuracy ±0.5% of full scale
Communications FK communications (DC power supply type)
Number of detector connectable: Up to 10 sets
Output renewal cycle 20ms, 56ms, 168mm (depended on computation modes)
Display LCD 16 digits
Settings Functional keys or FK communications
Calibration curve Linear to cubic expressions
Number of calibration curve 99 curves
Calibration curve creation Regression computation
Calibration curve correction Zero and span correction
External temperature correction Pt100ohm (option)
Computation Ratio/logarithm computation/multiregression computation, sand surface water ratio computation
Smoothing 0 to 99 seconds (0.1 second increment)
Calibration By a choking plate
Settings Settings including calibration curve No., calibration, preset output and hold
Self-diagnosis LCD display or output by FK communications for abnormal motor rotation, lamp failure, abnormal communications, high/low ambient temperature, sample low reflection, etc.
Working temperature 0 to 50°C
Power voltage 24VDC (20 to 28VDC)
Power consumption Maximum 25W
Allowable vibration Less than 3G
Connection Power/signal wires, detachable 8P terminal connection
Fiber optics type, Fiber optic connector connection
Case Aluminum drip-proof structure (conforming to IEC529 IP65)
Installation Bolt-hanging system, 4 pieces of M8 bolt
Weight About 3.7kg, about 8.5kg for explosion-proof type
  • Trace moisture meter detector unit
Optics Mirror system
Constituent 1constituent
Measuring distance 200mm (160 to 300mm possible)
Measuring area 30 x 30mm (at 200mm measuring distance)
Input/output terminals 1 analog output + 2 alarm outputs or 1 analog output +1 Pt100 input (for sample temperature correction)
Other specifications are same as standard specifications of constituent meters.
  • Explosion-proof type detector unit
Optics Mirror system, Fiber optics system
Wavelength 2, 3, 5, 6-wavelength
Constituent 1, 2, 3 constituents
Measuring distance Mirror system … 250mm (200 to 350mm possible)
Fiber optics system … Shown separately
Measuring area Mirror system … 50 x 50mm (at 250mm measuring distance)
Fiber optics system … Shown separately
Case Aluminum drip-proof structure (conforming to IEC529 IP65)
Weight About 8.5kg
Marking of explosion-proof construction ExdIIBT5
Exclusive bus cable for explosion proof IR-WVB
Input/output connection cable for explosion proof IR-WVRT
Other specifications are same as standard specifications of constituent meters.
  • Ultra-thin thickness meter detector unit
Optics Mirror system
Measuring area phi25mm (at 300mm measuring distance)
Special spot diameter 50 x 100mm, etc.
Power voltage 24VDC (20 to 28VDC)
Weight About 4.7kg
Other specifications are same as standard specifications of constituent meters.
  • Projection/ receiving separated type thickness meter detector unit
Measuring system Projection/ receiving separated type near-infrared absorption system
Measuring range 10 to 3000µm
Measuring distance 50mm (between projection unit and receiving unit)
Measuring diameter About phi6mm (phi25mm possible)
Output terminals 1 analog output
Power voltage 24VDC (20 to 28VDC)
Power consumption Maximum 30W
Connection Power/signal wires, detachable 8P terminal connection
16-core cable between projection unit and receiving unit
Weight About 3.5kg (projection unit/receiving unit)
Other specifications are same as standard specifications of constituent meters.

Fiber optics

  • Reflection type fiber optics (Biforked bundle fiber optics type)
Measuring diameter and distance phi20/15mm to phi50/50mm
Fiber optics length Standard 1.5m (maximum 10m)
Minimum vending radius R100mm
Working temperature 0 to 150°C
Fiber sheath Stainless steel snake tube
Purge air flow 5 to 20Nl/min
Accessories Horizontal mounting holder, flange holder
  • Reflection type fiber optics with lens (Biforked bundle fiber optics type)
Measuring diameter and distance phi20/25mm to phi40/100mm
Air purge function Not provided
Other specifications are same as specifications of reflection type fiber optics.
  • Accessories
Models Types
IR-WEA Air purge hood
IR-WEB Output checking plate
IR-WFK Air cooling fin
IR-WFW Water cooling plate
IR-WFP Mounting plate
  • Software package
Software Models Contents
Engineering software FK7001 For operation and setting of detector unit through FK bus
Software for exporting to Excel FK7201 For data acquisition, storage trend display and calibration curve creation of constituent meters (up to 10 detector units connectable)
Environment:
OS: Windows95/98/ME/NT4.0/2000/XP
I/F module: FK operational computer interface FK6501
Software for exporting to Excel: Microsoft Excel97/2000/2002 and FK engineering software FK7001 are required in addition to the above.

Page Top


BACK