Infrared
Radiation Thermometers
On-Line
Type
|
 |
 |
|
 |
|
IR-CA Series
High-speed Compact Radiation Thermometers |
| |
|
| The IR-CA series is a non-contact
and high-speed (3ms) compact radiation thermometer covering
medium and high temperature range by using the detecting
element of InGaAs or Si. The thermometer with converting
function has a digital display of temperature, parameter
setting keys, and a direct viewfinder. The thermometer
can be placed at any place due to a large distance factor.
|
 |
Features
- Functional digital display
of temperature and parameter setting keys
- Free installation by
a direct viewfinder and a long distance factor
- High-speed response
of 3ms
- Wide range
- High accurate type with
a short spectral response
- No moving part like
as chopper guarantees long reliability
- Various accessories
for every kinds of installation atmosphere
|
Models
| Models |
Type |
| IR-CAI |
For medium
temperature |
| IR-CAS |
For high
temperature |
| IR-CAQ |
High functional
type |
| IR-CAB |
Low temperature/long
spectral response |
| IR-CAE |
Low short/short
spectral response |
| IR-CAP |
For low/medium
temperature |
| IR-CAW |
Ultra-wide
type |
| IR-CAN |
For polyester
film |
| IR-CAM |
For polyethylene
film |
| IR-CAR |
For furnace
inside |
| IR-CAG |
For glass |
| IR-CAT |
For silicone |
| IR-CAU |
For gallium
arsenide |
Specifications
| Model |
For
medium temp
IR-CAI |
For
high temp
IR-CAS |
High
functional type
IR-CAQ |
| Measuring
system |
Single-color
type |
Two-color/single
color selectable |
| Element |
InGaAs |
Si |
InGaAs/InGaAs/Si |
| Spectral
response |
1.55µm |
0.9µm |
1.55/1.35/0.9µm |
| Measuring
range |
300 -
1600°C
200 - 1000°C
400 - 2000°C |
600 -
3000°C
500 - 2000°C
700 - 3500°C |
400 -
3100°C
350 - 2000°C
500 - 3500°C |
| Accuracy
rating *1 |
Less
than 1000°C
±5°C
1000°C to lower than 1500°C
±0.5%
of measured value
1500°C to lower than 2000°C
±1%
of measured value
2000°C or higher
±2% of measured
value |
| Resolution |
0.5°C |
1.0°C |
| Response
time |
3ms |
20ms |
| Optics |
Focusable
type |
| Measuring
diameter |
Measuring
distance/distance factor
Distance factor:
200, 300 (option), 50 (option)
200 with field diaphragm 10
(option)
300 with field diaphragm 10
(option) |
| Sighting |
Direct
viewfinder or laser spot (option) |
| Power
consumption |
Maximum
3VA |
*1: At  1.0
and reference operation condition
Reference condition: 23 ± 5°C, 35 to 75%RH
| Model |
Low
temp/
long wavelength type
IR-CAB |
Low
temp/
short wavelength type
IR-CAE |
For
low/
medium temp
IR-CAP |
Ultra-wide
type
IR-CAW |
| Measuring system |
Wide-band radiation thermometer |
Narrow-band radiation thermometer |
Wide-band/narrow-band radiation thermometer |
| Element |
Pyroelectric |
PbSe |
PbS |
TP/InGaAs/Si |
| Spectral
response |
8 - 13µm |
4µm |
2µm |
8 - 13/1.55/0.9µm |
| Measuring
range |
-50 - 100°C |
20 - 1000°C |
30 - 200°C |
100 - 500°C |
80 - 250°C
150 - 450°C
200 - 800°C |
200 - 3000°C |
| Accuracy
rating *1 |
±0.8°C |
Less than 200°C
±2°C
200°C or higher
±1% of measured
value |
±2°C |
±3°C |
Less than 500°C
±3°C
500°C or higher
±5°C |
Less than 1000°C
±5°C
1000°C to lower than 1500°C
±0.5%
of measured value
1500°C to lower than 2000°C
±1%
of measured value
2000°C or higher
±2% of measured
value |
| Resolution |
0.1°C |
1°C |
0.1°C |
1°C |
| Response
time |
2s |
0.2s |
0.02s |
0.1s |
| Optics |
Fixed
focus type |
Focusable
type |
Fixed focus type |
| Measuring
diameter |
37/1000mm, 20/1000mm
15/400mm, 10/500mm
8/200mm, 4/200mm
40/2000mm, 2/100mm |
37/1000mm
15/400mm
8/200mm |
Measuring distance/distance
factor
Distance factor: 200 |
Measuring distance/distance
factor
Distance factor: 50, 200, 300 |
29/1000mm |
| Sighting |
Laser
spot, without direct viewfinder |
Direct
viewfinder |
| Power
consumption |
Maximum
5VA |
Maximum
10VA |
Maximum 3VA |
*1: At  1.0
and reference operation condition
Reference condition: 23 ± 5°C, 35 to 75%RH
| Model |
For polyester film IR-CAN |
For polyethylene film IR-CAM |
For furnace inside IR-CAR |
For glass IR-CAG |
For silicone IR-CAT |
For gallium arsenide IR-CAU |
| Measuring system |
Narrow-band radiation thermometer |
| Element |
Pyroelectric |
PbSe |
MCT |
Si |
| Spectral
response |
8µm |
3.43µm |
3.8µm |
5µm |
0.6 - 0.96µm |
0.6 - 0.9µm |
| Measuring
range |
0 - 300°C |
30 - 300°C |
350 - 1100°C
450 - 1300°C
500 - 1500°C |
100 - 800°C
200 - 1800°C
400 - 2800°C |
400 - 800°C
500 - 1000°C
600 - 1200°C |
400 - 800°C
500 - 1000°C |
| Accuracy
rating *1 |
Less
than 200°C
±2°C
200°C or higher
±1% of measured
value |
Less than 1000°C
±5°C
1000°C or higher
±0.5% of measured
value |
Less than 1000°C
±5°C
1000°C to lower than 1500°C
±0.5%
of measured value
1500°C to lower than 2000°C
±1%
of measured value
2000°C or higher
±2% of measured
value |
Less than
600°C
±3°C
600°C or higher
±0.5% of measured
value |
| Resolution |
1°C |
0.5°C |
| Response
time |
1s |
0.02s |
0.1s |
0.04s |
| Optics |
Fixed
focus type |
Focusable
type |
| Measuring
diameter |
37/1000mm
15/400mm
8/200mm |
Measuring distance/distance
factor
Distance factor: 100, 200 |
Measuring distance/distance
factor
Distance factor: 50, 100, 200 |
Measuring
distance/distance factor
Distance factor: 100, 200 |
| Sighting |
Laser
spot, without direct viewfinder |
Direct
viewfinder |
| Power
consumption |
Maximum 5VA |
Maximum
10VA |
Common Specifications
| Emissivity
compensation |
Emissivity
setting range
1.9999 to 0.050 |
| Signal
modulation |
DELAY,
PEAK |
| Display |
Temperature,
parameters, 4 digits |
| Analog
output |
4 to 20mADC,
Isolate output |
| Working
temperature |
0 to 50°C |
| Power
supply |
24VDC
(Allowable voltage fluctuation range
22 to
28VDC) |
| Weight |
About
1.3kg |
| CE |
EN61326
+ A1 Emission Class A Immunity Annex A (complying
with connector connection only) |
*PDF data sheet is available at
"Download Center".
Page Top
BACK
|
|
|
|