Infrared
Radiation Thermometers
On-Line
Type
|
 |
 |
|
 |
|
IR-FA Series |
| |
|
The IR-FA series is a fiber
optic radiation thermometer featuring multi-function and
high-speed response. Three types, single-color type for
low temperature, single-color type for medium/high temperature
and two-color type, are available. Parameter setting function,
signal modulation and analog output are built in. Various
options including lens assembly with finder, laser-spotting
function, analog input and communications interface are
available.
|
 |
Features
- Low temperature type
with high-speed response (10ms) and short wavelength
enables to measure objects with low emissivity like
as metals.
- Medium/high temperature
type with high accuracy (±0.5 to 1% for 1000
to 2000°C) and high-speed response (10ms) is for
various temperature measurement fields.
- Two-color type is stable
with less effectiveness of smoke, vapor, dust and
lack-of-view.
- Digital temperature
display and settings of parameters by key operation
- Small and lightweight
(90 x 90 x 60mm, about 250g), DIN-rail installation
- Temperature measurement
of inductive heating object, measurement in explosion-proof
environment, measurement in vacuum equipment and other
measurement are possible.
- By using heat-resistive
fiber optic, measurement the environment at 150°C
is possible without any water-cooling.
- By signal modulation
function, the stable temperature measurement is possible.
- Laser function for easy
spotting of measuring point (option)
- A lens assembly with
finder for spotting of measuring point with eyes is
available from accessories.
- By the communications
interface (RS-485), the data logging and parameters
setting are enabled from a personal computer.
- Emissivity setting (emissivity
ratio for two-color type) by analog input or automatic
emissivity computation function is selectable. (option)
- CE conformance
|
Models
[Thermometers]
| Elements |
Models |
| Cooling type PbS |
IR-FAC   |
| InGaAs |
IR-FAI   |
| Si |
IR-FAS   |
| Hybrid element |
IR-FAQH  |
[Lens assembly]
Core diameter
400µm
| Measuring
diameter/measuring distance |
Models |
1mm
at 100mm |
IR-FL0   |
12mm
at 1000mm |
IR-FL1   |
5mm
at 500mm |
IR-FL2   |
2mm
at 200mm |
IR-FL3  |
4mm
at 200mm |
IR-FL4  |
5mm
at 150mm |
IR-FL5  |
20mm
at 600mm |
IR-FL6  |
8mm
at 1000mm |
IR-FL8  |
Core diameter
800µm (for low temperature)
| Measuring
diameter/measuring distance |
Models |
10mm
at 150mm |
IR-FL5   |
40mm
at 600mm |
IR-FL6   |
13mm
at 200mm |
IR-FL7   |
[Power unit]
| Name |
Model |
| Power unit |
IR-ZFEP |
Standard Ranges
| Types |
Elements |
Measuring
ranges |
Lens
assemblies |
| Single-color |
Cooling
type PbS |
70 to 250°C |
IR-FL5 ,IR-FL6 ,IR-FL7 core
dia 800µm) |
| 100 to 300°C |
| 250 to 800°C |
IR-FL0 ,IR-FL1 ,IR-FL2 ,IR-FL3 ,IR-FL4 (core
dia 400µm) |
| 150 to 500°C |
IR-FL5 ,IR-FL6 (core
dia 400µm) |
| 250 to 800°C |
| InGaAs |
150 to 450°C |
IR-FL5,IR-FL6 |
| 200 to 700°C |
| 250 to 1000°C |
| 300 to 1300°C |
| 250 to 1000°C |
IR-FL0,IR-FL1,IR-FL2,IR-FL3,IR-FL4,IR-FL8 |
| 300 to 1300°C |
| 350 to 1600°C |
| Si |
400 to 900°C |
IR-FL5,IR-FL6 |
| 500 to 1200°C |
| 600 to 1800°C |
| 700 to 2400°C |
| 600 to 1800°C |
IR-FL0,IR-FL1,IR-FL2,IR-FL3,IR-FL4,IR-FL8 |
| 700 to 2400°C |
| 800 to 3000°C |
| 2-color
type |
Hybrid
element |
500 to 1000°C |
IR-FL5,IR-FL6 |
| 600 to 1500°C |
| 700 to 2000°C |
| 1000 to 2400°C |
| 600 to 1500°C |
IR-FL0,IR-FL1,IR-FL2,IR-FL3,IR-FL4,IR-FL8 |
| 700 to 2000°C |
| 800 to 2400°C |
| 1000 to 3000°C |
Specifications
[Thermometer]
| Model |
IR-FAC |
IR-FAI |
IR-FAS |
IR-FAQH |
| Measuring
system |
Single
color type |
Two- color
type |
| Element |
PbS (cooling
type) |
InGaAs |
Si |
Hybrid
Element |
| Wavelength |
2.0µm |
1.55µm |
0.9µm |
0.9 /
1.55µm |
| Accuracy
ratings |
70°C
to 300°C: ±4°C
300°C to 500°C: ±5°C
Higher than 500°C: ±1.0 % of measured
value |
Lower
than 1000°C: ±5°C
1000°C to 1500°C: ±0.5% of measured
value
1000°C to 1500°C: ±0.5% of measured
value
1500°C to 2000°C: ±1.0% of measured
value
Higher than 2000°C: ±2% of measured value |
| Resolution |
70°C
to 100°C: About 3°C
100°C to 200°C: About 3°C
Higher than 200°C: About 0.5°C |
0.5°C |
1.0°C |
| Response
time |
0.01 second |
0.04 second |
| Emissivity
(ratio) compensation |
1.999
to 0.050 (*1) |
Emissivity:
1.999 to 0.050 |
Emissivity
ratio: 1.999 to 0.050 |
| Signal
modulation |
DELAY,PEAK |
| Display |
LCD 4-digit
(Temperature display, parameter display), °C/°F
(key selection) |
| Analog
output |
4 to 20mADC
isolated output (load resistance: 500
or lower) |
| Contact
output |
2-point |
1-point |
| Contact
input |
1-point,
peak hold reset or sample hold, Dry contact or open
collector |
| Parameter
setting by keys |
Operator mode and Engineering mode |
| Working
temperature |
5 to 40°C |
0 to 50°C |
| Rated
power supply |
24VDC
(Allowable voltage fluctuation range: 22 to 28VDC),
Recommended power unit: IR-ZFEP |
| Power
consumption |
Maximum
15VA |
Maximum
3VA |
| Mounting |
DIN rail
mounting or wall mounting |
| Outside
Dimensions, Weight |
W140 x
H110 x D65mm About 1.0kg |
W90 x
H90 x D60mm. About 250g |
| CE |
EN61326
+ A1
Emission: Class A
Immunity: Table A.1 |
| Stability
under EMC test environment |
±10°C |
±10°C
or ±1% of measuring range, whichever is larger |
| Condition
of CE conformance |
See Note *2. |
|
*1: The effective compensation
range is e = 1.0 to 0.8 for the measured temperature of
70 to 80°C and
= 1.0 to 0.6 for 80 to 90°C.
*2: Condition of CE conformance for IR-FAC
(1) The connection cable is up to 30m.
(2) An individual DC power unit for thermometer is required.
(Recommended power unit: IR-ZFEP)
(3) All of the thermometer, the connection cable, and
a DC power unit should be used indoors.
[Fiber optics]
| |
Core
400µm |
Core
800µm (IR-FAC) |
| Fiber
optics |
Single core quartz |
| Sheath |
Without
metallic protective tube (Heat resistive sheath/glass
wool braid)
With metallic protective tube (Heat resistive sheath/glass
wool braid + SUS flexible tube) |
| Working
temp. |
0-150°C |
0-50°C at measured
value of 70-120°C 0-80°C at measured value
of 120°C or higher |
| Length |
4m (standard) for
IR-FAC (max.5m)
Specify 2-50m for IR-FAI, S or QH |
2m |
| Allowable
bending |
R100mm |
R150mm |
*PDF data sheet is available at
"Download Center".
Page Top
BACK
|
|
|
|