Traceability
Backbody Furnace
|
 |
 |
|
 |
|

|
NRLM
Type Comparison Blackbody Furnaces IR-R24, IR-R26 High Temperature Comparison Blackbody Furnaces IR-R27
|
|
|
|
| NRLM type comparison
blackbody furnaces, which have been jointly developed
with the National Research Laboratory of Metrology, are
large opening blackbody furnaces for calibration of radiation
thermometers, and a high temperature comparison blackbody
furnace uses a standard radiation thermometer for comparison
calibration.NRLM Type Comparison Blackbody Furnace |
 |
Features(IR-R24,
IR-R26)
- The cavity is profiled
as a double cone and the high emissivity is realized
with a large opening.
- For low temperature,
a platinum resistance thermometer (sold separately)
is used as a standard thermometer.
- For medium temperature,
a 0.9µm standard radiation thermometer (sold
separately) is used for comparison calibration.
- Long-term stability
|
|
Models
|
Models
|
Types
|
| IR-R24
|
NRLM
type comparison blackbody furnace |
For low
temperature |
| IR-R26
|
For medium
temperature |
| IR-R27
|
High
temperature comparison blackbody furnace |
Specifications
|
Model
|
NRLM
type comparison blackbody furnace
|
High
temperature comparison blackbody
furnace
|
|
For
low temperature
IR-R24
|
For
medium temperature
IR-R26
|
IR-R27
|
|
Temperature
range
|
50 to
450°C
|
200
to 1050°C
|
800
to 1450°C
|
|
Maximum
temperature
|
500°C
|
1100°C
|
1500°C
|
|
Maximum
temperature rising time
|
About
2 hours
|
About
4hours
|
About
2 hours
|
|
Cavity
opening diameter
|
60mm
|
50mm
|
30mm
|
|
Cavity
emissivity
|
0.993
|
0.997
|
0.99
|
|
Cavity
section shape
|
Double
cone type
|
Cylindrical
cone type
|
|
Temperature
stability
|
±
0.15K/30 minutes
|
±
0.25K/30 minutes
|
±
0.5K/30 minutes
|
|
Power
supply
|
100VAC,
Single phase
|
200VAC,
Single phase
|
|
Power
consumption
|
Maximum
500VA
|
Maximum
1200VA
|
Maximum
2500VA
|
|
Weight
|
About
30kg
|
About
32kg
|
About
36kg
|
Page Top
BACK
|
|
|
|