Measuring
points
|
Dot-printing
type |
6
points |
| Pen-writing
type |
1-pen,
2-pen or 3-pen |
| Input
signals |
DC
voltage |
±7mV,
±14mV, ±25mV, ±70mV, ±5V |
| DC
current |
Applicable
by adding a shunt resistor (100 ,
250 ) |
| Thermocouples
|
B,
R, S, K, E, J, T, N, Ni-NiMo, AuFe-Cr, PR5-20, PR20-40,
WWRe5-26, WWRe0-26,
Platinel, U, L |
| Resistance
thermometers |
Pt100,
JPt100, Pt50, Pt-Co
|
| Accuracy
ratings |
Digital
displays and printing
|
Thermocouples,
resistance thermometers |
±0.3%
of the scale range ± 1 digit or ±
1°C, whichever is larger |
| DC
voltage |
±0.2%
of the measuring range ± 1 digit |
| Analog
indication |
±0.5%
of the scale range |
|
Note: For thermocouple inputs, the
reference junction compensating accuracy is
excluded. |
| Measuring
cycles |
Dot-printing
type |
About 5 seconds/6-point |
| Pen-writing
type |
About 125ms |
| Reference
junction compensating accuracy |
K,
E, J, T, N, Platinel: Maximum ±0.5°C
R, S, Ni-NiMo, AuFe-CR, WWRe5-26, WWRe0-26, U, L:
Maximum ±1.0°C |
| Dot-printing
cycle |
About
5 seconds/point (when the chart speed is faster
than 6mm/h.) |
| Printing
systems |
Dot-printing
type |
Trace
and digital printings: Wire dot system 6-color ribbon |
| Pen-writing
type |
Trace
printing: Cartridge pen
Digital printing: Plotter pen |
| Printing
colors |
Dot-printing
type |
Trace
printing |
Red,
black, blue, green, brown, purple |
| Digital
printing |
Periodic
digital data printing: Repetition of 6 colors
Digital data printing: Repetition of 6 colors
Date and time printing: Black
Scale and point No.: Same
colors as dot printing colors
Chart speed: Black
Alarm activation/reset: Red
(option)
Programming change mark: Black |
|
Pen-writing type |
Trace
printing |
1st
pen: red, 2nd pen: green, 3rd
pen: blue |
| Digital
printing |
Purple
(Periodic digital data printing, digital data printing,
date/time printing,
Scale/chart speed/programming change mark/optional
alarm activation/reset
printing) |
| Chart |
Fan-fold type 114mm total width, 10m total length,
100mm effective printing width |
| Chart
speed |
1 to 1500mm/h for dot-printing type
1 to 599mm/h, 10 to 200mm/min for pen-writing type |
| Balancing
speed |
About
2 seconds (pen-writing type) |
| Periodic
digital data printing |
Time, point No., measured values, and engineering
units are printed on trace
printing at every specified time interval. |
| Digital
data printing |
Time, point No., measured values, and units are
printed at the request time by interrupting
trace printing. |
| Alarm
activation and reset printing |
Time, point No., alarm type and level are printed
on the right side of thechart
when an alarm activates.
Time, point No., and level
are printed on the right side of the chart when
the alarm was reset.
* In case
alarms are specified on ordering. |
| Display
items |
Multi-point
sequential display, 1-point continuous display,
and clock display are switched. |
| Alarm
type |
Absolute
alarms, differential alarms, rate-of-change alarms
(to be specified on ordering) |
| Alarm
setting |
Individual setting on each point, Up to 4 alarm
levels/point) (to be specified on ordering) |
| Rated
supply voltage |
90
to 120VAC, or 180 to 240VAC (to be specified) |
| Rated
supply frequency |
50Hz/60Hz
(selectable by DIP switches) |
| Scale
plates |
Dot-printing
type |
Up to triple scale, minimum 80 equal divisions |
| Pen-writing
type |
Up
to dual scale, minimum 80 equal divisions |
| Maximum
power consumption |
Dot-printing
type |
45VA |
| Pen-writing
type |
1-pen:
30VA, 2-pen: 35VA, 3-pen: 40VA |
| Working
temperature range |
0
to 50°C |
| Working
humidity range |
20
to 80%RH |
| Weight |
Dot-printing
type |
About 2.6kg |
| Pen-writing
type |
1-pen:
about 3.0kg, 2-pen: about 4.0kg, 3-pen: about 4.5kg |
| CE
(option) |
EN61326
+ A1 Class A, EN61000-3-2 + A14