Short Stack
type Fuel Cell Evaluation Test Equipment |
| |
|
The
equipment is PEFC type short stack evaluation test equipment
consisted of gas supply line, moisturizing line, cooling
line, cell performance line and safety line. |
|
Features
- Standalone and compact
equipment
- Mixed gas with optional
mixing ratio can be produced by setting total volume
and mixing ratio of each gas. Cell is protected by
over-flow protection function.
- Gas utilizing ratio
constant control. Tafel measurement (IV characteristics),
Resistance measurement (option)
- Stable humidification
control, Humidification unit water temperature (+30°C
to 90°C), Long term stable operation by humidification
unit self-cleaning system, High-speed humidification
response (within 3 min - option)
- High-speed response
cooling system
- Resistance measurement
at conduction of fuel cell (option)
- Various options including
pressure test, humidity measurement, etc.
- Automatic operation
by exclusive control software (CISAS-EX) for controlling
of parameters such as gas flow, bubbler temperature,
pipe temperature control, load current, etc. for evaluation
tests.
|
Specifications
Gas
supply line
| Gas |
Maximum
flow |
Control
range |
| Hydrogen
(H2) |
30NL to
min |
0.6 to
30NL/min |
| Carbon
dioxide (CO2) |
15NL/min
|
0.3 to15NL/min |
| Mixed
gas |
1NL/min |
0.02 to
1NL/min |
| Air |
100NL/min |
3.0 to
100NL/min |
Purge
nitrogen
Fuel side
Air side |
About
10NL/min each |
No control |
Flow setting accuracy: ±1%
of F.S.
Flow control range: 2% of
maximum flow to maximum flow
Repetition accuracy: Within
±0.25% (indicated value)
MFC response time: Within
1 second
Moisturizing
control line
Moisturizing range: Within
room temperature+30°C to 90°C±1°C (at
stabilized)
Moisturizing capacity: 95%RH
or more
Material: SUS316
Temperature drop at water
supply: Within 1°C
Water supply: Automatic power
supply (Ion exchanged water of 6M
or more)
Moisturizing tank automatic
cleaning system: Built-in
Stack
cooling line
Cooling water temperature:
+20°C to 90°C, ±1°C
Cooling water flow: 1 to 5L/min,
±1% flow meter range
| Control system: |
Inlet temperature/flow
constant control
Inlet temperature/temperature difference constant
control |
Cooling capacity: 1.5KW
Running pressure control
line (option)
Pressure control range: maximum
500kPa, within ±5kPa
Pressure response: Changing
by test gas flow/equipment capacity
Operation control software
Stable software being used
in various industrial applications
Automatic forwarding of testing
items
Excellent performance to noise
Reduction of CPU load
Exclusive software part for
fuel cell
The software changes process conditions by sequencing
numbers of sheets writing test. After judgment of stability,
it automatically collect data.
(Electronic load unit control)
(1) Initial activation automatic
operation
(2) Load current sweep test=Tafel
test
Electronic load (several
hundred W to several hundred kW)
As a bias power supply corresponding
to 0V, it can be used in a single cell.
3 ranges of current settings,
high accuracy of ±0.3% of F.S.
For wide frequency from 5mHz
to 30kHz
For resistance measurement
of IR drop, frequency fixed alternative method, frequency
sweep alternative method
| Items |
Specifications |
| Load mode |
Constant
current/constant resistance/constant power |
| Discharge
voltage range |
0 to130V |
| Maximum
current |
0 to 1,560A |
| Maximum
load power |
0 to 100kW
|
| Current
setting |
3-range,
±0.3% of F.S. |
| Constant
voltage setting |
0 to 130V/0.01V,
±0.2% of F.S. |
| Constant
power setting |
3-range,
±0.8% of F.S. |
| Voltage
measurement |
1-range,
±0.1% of F.S. |
| Current
measurement |
2-range,
±0.3% of F.S. |
| Resistance
measurement |
3-range,10/100/1000m |
Page Top
BACK
|
|
|
|