Models
| Models |
Types |
| Low temperature |
IR-SAB |
| Medium temperature |
IR-SAI |
| High temperature |
IR-SAS |
| Two-color type |
IR-SAH |
Connectivity
Specifications
|
Low temperature |
Medium temperature |
High temperature |
2-color type |
| Model |
IR-SAB |
IR-SAI |
IR-SAS |
IR-SAH |
| Element |
PE |
InGaAs |
Si |
Si/InGaAs |
| Measuring wavelength |
8 to 14µm |
1.55µm |
0.9µm |
0.9/1.55µm |
| Measuring range |
0 to 1000°C |
300 to 1600°C |
600 to 2500°C |
900 to 2500°C |
| Accuracy rating |
200°C or less: ±2°C
200°C or
more: ±1% of measuring
value |
1000°C or less: ±0.2% of measuring
value ±2°C
1000 to 1500°C:
±0.4% of measuring
value
1500°C
or more: ±0.5% of measuring
value |
1500°C or less: ±0.5% of measuring
value
1500°C
or more: ±0.6% of measuring
value |
| Resolution |
0.5°C |
1°C |
| Response time (95%) |
0.2s |
0.002s |
0.01s |
| Optics |
Fixed focus lens type |
| Sighting |
Laser unit |
Telescope or laser pointer |
| Parameter setup |
Through RS485 |
Emissivity adjustment
(2-color: emissivity
ratio) |
1.9999 to 0.200 |
1.9999 to 0.050 |
1.250 to 0.750 |
| Analog output |
4 to 20mA DC isolated output
Output scaling:
Programmable in measuring
range |
| Working temperature |
0 to 50 °C |
0 to 90°C |
| Power supply |
24V DC±10% |
| Power consumption |
Approx. 5.0VA |
Approx. 2.4VA |
| Protection |
IP67 |
| Weight |
Approx. 0.7Kg |
| Standards |
CE approval, RoHS directive |
Measuring diameter & distance
Setting display unit IR-GZ
The IR-GZ is combined with the IR-SA for
enabling parameters setup, data display and
24V DC power supply to the IR-SA. |
Specifications
| Emissivity (ratio) setting |
1.999 to 0.000 |
| Thermometer input |
RS485 |
| Signal modulation |
DELAY,PEAK |
| Display |
Temperature, Thermometer number being connected,
Status display |
| Analog output |
Output 1: 4 to 20mA DC (IR-GZ output)
Output 2: 4 to 20mA DC (IR-SA) |
| Event output |
2 points |
| Communications interface |
RS232C (Optional) or RS485 (Optional) |
| Power supply to IR-SA |
24V DC |
| Power supply |
100 to 240V AC, 50/60Hz |
| Power consumption |
Maximum 20VA |
| Working temperature |
-10 to 50°C |
| Working humidity |
20 to 90%RH (No dew condensation) |
| Weight |
Approx 0.5Kg
CE marking EMC
directive EN61326+A1
Low voltageENN61010-1+A2
Overvoltage
category II,Pollution level 2 |
*PDF data sheet is available at
"Download Center".
Page Top