IR-CZ Series Radiation Thermometers

IR-CA series is self-contained type, stationary radiation thermometer equipped with digital temperature display, and measured value external output.

  • Highly accurate measurement at high temperature range is achieved by tracing from a metal-carbon eutectic point.
  • It is capable of measurement from 500°C as a single color type.
  • Stability at high temperature continuous measurement and resolution from low temperature range are improved.
*Refer to the page of IR-CA series for setting display unit IR-GZA used for power supply, display and parameter setting etc.

 

Models

Types Measuring Range Models
For high temperature 2-color type radiation thermometer 900 to 3500°C IR-CZQH7
For medium temperature 2-color type radiation thermometer 400 to 2000°C IR-CZQW2

Specifications

Types For high temperature 2-color type radiation thermometer For medium temperature 2-color type radiation thermometer
Models IR-CZQH7 IR-CZQW2
Measuring Wavelength 0.9/1.55μm 0.9/1.35/1.55μm
Detecting element Si/InGaAs Si/InGaAs/InGaAs
Accuracy Rating Below 1500°C…±5°C
1000°C or more, below 1500°C…±0.5% of measured value
1500°C or more, below 2500°C…±0.6% of measured value
2500°C or more…±1.0% of measured value
Below 1000°C…±5°C
1000°C or more, below 1500°C…±0.5% of measured value
1500°C or more…±0.6% of measured value
Response Time 2 to 15ms
Distance Factor 200 with field spot φ10mm 200
Focusing System for Spot Finder focusing
Analog Output 4 to 20mA DC Isolated output
Load resistance…750Ω or less
Communication Interface RS-485 transmits measured data, transmits and receives various parameters
Power Supply 24V DC (Allowable voltage fluctuation range・・・22 to 28V)
Power Consumption Max. 2.4VA
Connection Dedicated connector connection, terminal connection