IR-BA Series High-speed Compact Radiation Thermometers

Data_Sheet
The IR-BA series is a non-contact and high-speed compact radiation thermometer. The card sized thermometer with built-in signal converting function of 4-20mA output can easily be installed in narrow places, production lines and small equipment.
A standard speed type and a high-speed type (50ms/95%) are respectively available in standard type and small-target/short-distance type and small-target/long-distance type, and special application types for glass, lamp-anneal, film, and burning gas..

Features

  • High-speed response
    Standard: 100ms, High-speed: 50ms
  • Compact & light-weight
    W32 x H56 x D100 mm, 220g
  • Signal conversion built-in
  • Emissivity setting by digital switch
  • Peripheral units including:
    Indicator with power supply, converter with reflection correction, etc.
  • CE
    Excluding special application types (IR-BAX)

Models

Models Types Range Response D/L Accuracy Wavelength
IR-BAT1A General type Standard 0-300°C 100ms Φ40/500mm Less than 300°C:
±3°C
300°C or higher:
±1% of measured value
(at Φ = 1.0 and reference operating condition)
8 to 14µm
IR-BAT1B 0-600°C
IR-BAT2A Small target/
Short distance
0-300°C Φ5/80mm
IR-BATSA Small target/ 0-300°C Φ40/1000mm
IR-BATSB Long distance 0-600°C
IR-BAT1M High-speed type Standard 0-300°C 50ms Φ40/500mm
IR-BAT1P 0-600°C
IR-BAT2M Small target/
Short distance
0-300°C Φ5/80mm
IR-BATSM Small target/
Long distance
0-300°C Φ40/1000mm
IR-BATSP 0-600°C
IR-BAXG1 Application type Glass 100-400°C 1s Φ20/150mm ±4°C 4.6 to 5.2µm
IR-BAXL1 Lamp anneal 400-1300°C 1s Φ40/500mm 1% 3.43µm half-band width 120nm
IR-BAXF1 Polyethylene 80-250°C 1s Φ40/500mm ±4°C
IR-BAXF2 Polyester 50-150°C 1s Φ40/500mm ±4°C 7.6 to 8.4µm
IR-BAXH1 Burning gas 500-1300°C 10s Φ100/1000mm 1.5% 4.3µm

Specifications

Element Thermopile
Resolution 0.2°C (0-300°C) general type
0.5°C (0-600°C) high-speed type
Emissivity 1.99 to 0.10 by digital switch
Analog output 4 to 20mADC (Load: less than 280Ω)
Working temperature 0 to 50°C
Power 12 to 24VDC
Current consumption 60mA or lower
Connecting cable Standard 2m 5m for IR-BAXH1
Water proofing JIS C-920 Drip-proof II, IP-X2
Weight About 220g
CE EN61326 Class A, EN61326 Annex A
Except special application types and extension cable type
Conditions for CE conformance 1. An individual DC power unit for thermometer is required.(Recommended power unit: IR-ZFEP)
2. All of the thermometer, the connection cable, and a DC power unit should be used indoors.
Stability ±10°C under the EMC test environment